Occupational Health and Safety Regulation, B.C. Reg. 296/97

Section 30.9

Airflow and containment monitoring

Cite as Occupational Health and Safety Regulation, B.C. Reg. 296/97, s. 30.9 · last amended September 1, 2027 · read from BC Laws on November 17, 2026

The text

(1) Face velocities over the operational face opening of a laboratory fume hood must be quantitatively measured and recorded. (2) The ability of a laboratory fume hood to (a) maintain an inward flow of air across the operational face opening, and (b) contain contaminants must be assessed and recorded using a smoke tube or other suitable qualitative method. (3) The actions described in subsections (1) and (2) must be performed (a) after the laboratory fume hood is installed and before it is used, (b) at least once in each 12 month period after installation, and (c) after any repair or maintenance that could affect the air flow of the hood. (4) If a laboratory fume hood is found to be operating with an average face velocity of less than 90% of the average face velocity required in section 30.8 (2), the employer must immediately take corrective action to bring the average face velocity within the required range of velocities. (5) Airflow in a laboratory fume hood must be monitored continuously if loss of airflow will result in risk to a worker. (6) A laboratory fume hood that is being installed must have an alarm capable of indicating when the average face velocity falls below the minimum average face velocity level required in section 30.8 (2) when the hood is in use.
Read the Occupational Health and Safety Regulation on BC Laws ↗

Made under Workers Compensation Act, RSBC 2019, c. 1.