Occupational Health and Safety Regulation, B.C. Reg. 296/97
Section 30.8
Fume hoods
Cite as Occupational Health and Safety Regulation, B.C. Reg. 296/97, s. 30.8 · last amended September 1, 2027 · read from BC Laws on November 17, 2026
The text
(1) A laboratory fume hood and its related ductwork must be designed, installed and maintained in accordance with the Industrial Ventilation, A Manual of Recommended Practice, published by the American Conference of Governmental Industrial Hygienists, as amended from time to time.
(2) A laboratory fume hood must
(a) be connected to a local exhaust ventilation system,
(b) provide average face velocities of 0.4 m/s (80 fpm) to 0.6 m/s (120 fpm) across the operational face opening,
(c) not have face velocities of less than 80% of the average face velocity required in paragraph (b) at any point across its operational face opening, and
(d) not have face velocities of more than 120% of the average face velocity required in paragraph (b) at any point across its operational face opening.
(2.1) A laboratory fume hood must have a sash that is positioned to protect the upper body and face of a worker working in the laboratory fume hood from accidental releases of the contents of the hood while allowing hand and arm access to equipment inside the hood.
(2.2) A laboratory fume hood with a movable sash must be clearly marked to identify the maximum size of the operational face opening that will maintain the average face velocities required in subsection (2) (b).
(2.3) The employer must ensure
(a) that before it is used, a commercially manufactured laboratory fume hood has been certified as being tested by the manufacturer, and
(b) following installation and before it is used, a custom built laboratory fume hood is tested on site by a qualified person.
(2.4) A laboratory fume hood tested under subsection (2.3) must demonstrate containment not greater than the control level of 0.05 ppm when tested under "as manufactured" test conditions in accordance with the methods described in ANSI/ASHRAE Standard 110-1995, Method of Testing Performance of Laboratory Fume Hoods.
(2.5) The installation of a laboratory fume hood must be certified by a professional engineer.
(3) A laboratory fume hood must be located to prevent cross drafts or other disruptive forces from lowering the air flow across the operational face opening to unacceptable levels.
(4) A laboratory fume hood and its ductwork must be constructed from materials compatible with its use.
(5) A laboratory fume hood that will be or is being used for working with
(a) radioactive material in amounts that exceed the exemption quantity specified by the Canadian Nuclear Safety Commission, or
(b) perchloric acid
must be clearly labelled with applicable restrictions on its use.
(6) A laboratory fume hood must not be used for storage of chemicals unless it is used exclusively for this purpose and is labelled with this limitation.
(7) Controls for the operation of a laboratory fume hood and its service fixtures must be
(a) located on the outside of the laboratory fume hood, and
(b) immediately accessible to the worker conducting work in the laboratory fume hood.
(8) Despite subsection (7), water taps may be located inside a laboratory fume hood if the main shutoff valve for the water is located outside the laboratory fume hood.
(9) Equipment being used in a laboratory fume hood must
(a) be kept at least 15 cm (6 in.) from the operational face opening of the laboratory fume hood, and
(b) not adversely affect airflow into the laboratory fume hood.
(10) Written procedures must be developed and implemented to ensure safe use and operation of a laboratory fume hood.
Read the Occupational Health and Safety Regulation on BC Laws ↗
Made under Workers Compensation Act, RSBC 2019, c. 1.