Multi-Sector Air Pollutants Regulations, SOR/2016-151
Section 83
Sequence of calibration error checks
Cite as Multi-Sector Air Pollutants Regulations, SOR/2016-151, s. 83 · current to January 19, 2026 · regulation last amended January 1, 2026
Statute Watch
Get an alert when a bill in Parliament would change this section, and every time it moves.
The text
(1) A sequence of calibration error checks consists of two calibration error checks on each of the electrochemical cells of an analyzer, one involving the introduction of a zero gas and the other involving the introduction of a span gas.
(2) A calibration error check on an electrochemical cell of an analyzer is conducted by introducing into the analyzer, at a constant flow rate, a calibration gas — namely, a zero gas or a span gas — for each of the gases referred to in subsection (3) to which that cell is designed to respond in order to determine the difference between
(a) the concentration of that gas in that calibration gas, as indicated by the reading of that cell’s response to it taken after the stabilization period for the calibration error check for that cell, and
(b) the certified concentration of that gas in that calibration gas.
(3) The zero gas and the span gas for the following gases to which a cell is designed to respond are
(a) for O2,
(i) a zero gas that consists of
(A) O2 in nitrogen, with a certified concentration of O2 of less than 0.25% of the concentration of O2 in its corresponding span gas,
(B) CO in nitrogen, or in air, with a certified concentration of O2 of less than 0.05%, or
(C) NO in nitrogen, with a certified concentration of O2 of less than 0.05%, and
(ii) a span gas that consists of
(A) ambient air in a well-ventilated area, with a presumed certified concentration of 20.9% O2, or
(B) O2 in nitrogen that has a certified concentration of O2 that is(I) within the measuring range, specified by the analyzer’s manufacturer, for the O2 cell, and(II) at least 5% and at most 20.9%;
(b) for CO,
(i) a zero gas that consists of
(A) ambient air in a well-ventilated area, with a presumed certified concentration of 0 ppm CO, or
(B) CO in nitrogen, or in air, with a certified concentration of CO of less than 0.25% of the concentration of CO in its corresponding span gas, and
(ii) a span gas that consists of CO in nitrogen, or in air, with a certified concentration of CO that is
(A) within the measuring range, specified by the analyzer’s manufacturer, for the CO cell, and
(B) at least 200 ppm and at most 400 ppm;
(c) for NO,
(i) a zero gas that consists of
(A) ambient air in a well-ventilated area, with a presumed certified concentration of 0 ppm NO, or
(B) NO in nitrogen, with a certified concentration of NO of less than 0.25% of the concentration of NO in its corresponding span gas, and
(ii) a span gas that consists of NO in nitrogen, with a certified concentration of NO that is
(A) within the measuring range, specified by the analyzer’s manufacturer, for the NO cell, and
(B) at least 100 ppm and at most 200 ppm; and
(d) for NO2,
(i) a zero gas that consists of
(A) ambient air in a well-ventilated area, with a presumed certified concentration of 0 ppm NO2, or
(B) NO2 in air with a certified concentration of NO2 of less than 0.25% of the concentration of NO2 in its corresponding span gas, and
(ii) a span gas that consists of NO2 in air, with a certified concentration of NO2 that is
(A) within the measuring range, specified by the analyzer’s manufacturer, for the NO2 cell, and
(B) at least 50 ppm and at most 100 ppm.
(4) A concentration of O2, CO, NO or NO2 in a calibration gas is certified if the manufacturer of the gas certifies that
(a) the concentration of that gas in the calibration gas is accurate within ± 2%; and
(b) the calibration gas was prepared and analyzed
(i) in accordance with section 2 of method entitled EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards, published in May 2012 and bearing the designation EPA/600/R-12/531, or
(ii) by means of a process for measuring that concentration and for verifying its accuracy in accordance with specifications and standards of the National Institute of Standards and Technology of the United States.
(5) The stabilization period for a sequence of calibration error checks on an analyzer is the longest period among the periods required for each electrochemical cell to have a stable response, following the introduction of a span gas into the analyzer, to the gas to which the cell is designed to respond. The duration of each of those periods must be recorded.
(6) The calibration error — for each calibration error check involving the introduction of the zero gas or of the span gas — must be at most
(a) for O2, 0.5%; and
(b) for CO, NO or NO2, the greater of
(i) 5% of the certified concentration of that gas in the span gas, and
(ii) 10 ppm.