X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543
Section 17
Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 17 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026
The text
(1) Every fluoroscopic x-ray machine shall be equipped with,
(a) an image intensification system that,
(i) includes a shielded protective barrier and shielding such that,
(A) the entire cross-section of the useful beam is intercepted within the protective barrier for any target-to-image distance, and
(B) the fluoroscopic x-ray tube is not capable of producing x-rays unless the shielding is in place to intercept the useful beam,
(ii) in the case of a mobile fluoroscopic x-ray machine, is an integral part of the machine or is interlocked in such a manner that its removal prevents x-rays from being produced;
(b) where it is a stationary machine, a means to prevent the x-ray tube from producing x-rays unless there is an image receptor in place to intercept the x-ray beam;
(c) an audible signal that,
(i) indicates completion of any preset time of use up to a maximum of five minutes, and
(ii) continues to sound until the timer is reset whenever x-rays are produced after the preset time of use has expired, or,
a timer circuit that will,
(iii) cut off the high tension voltage to the x-ray tube after a preset time of use up to a maximum of five minutes, and
(iv) continue to prevent fluoroscopy until the timer has been reset manually;
(d) electrical meters or other visual indicators on the control panel that will provide a continuous indication of current in milliamperes;
(e) a means to limit the target-to-skin distance to not less than,
(i) twenty-five centimetres for a mobile fluoroscopic machine,
(ii) thirty-eight centimetres for a stationary fluoroscopic machine, or
(iii) twenty centimetres for an image-intensified fluoroscopic machine used for special procedures that would not be possible at the minimum target-to-skin distance set out in subclause (ii);
(f) an x-ray beam limiting device that,
(i) allows the operator to adjust the dimensions of the x-ray field at the image receptor to a size smaller than the dimensions of the image receptor, and
(ii) aligns the x-ray field with the image receptor in such a manner that the x-ray field is always confined within the boundaries of the image receptor; and
(g) a shield of at least 0.25 millimetres lead equivalent at 100 kilovolts peak that intercepts scattered radiation originating in the patient that would otherwise reach the x-ray operator or other persons in the facility. R.R.O. 1990, Reg. 543, s. 17 (1); O. Reg. 87/22, s. 7 (1).
(2) Clause (1) (b) does not apply to special purpose x-ray tubes or image intensifiers that are constructed to have free and independent movement within an x-ray room. R.R.O. 1990, Reg. 543, s. 17 (2).
(3) Clause (1) (g) does not apply to a mobile fluoroscopic x-ray machine. R.R.O. 1990, Reg. 543, s. 17 (3).
(4) The exposure rate limit of a fluoroscopic x-ray machine that uses a zinc cadmium sulphide input phosphor or a phosphor of similar efficiency calculated where the centre of the useful beam enters the patient at the shortest target-to-skin distance specified for the machine shall not exceed,
(a) a maximum exposure rate of 12.5 roentgens per minute; and
(b) an entrance exposure rate of five roentgens per minute for an average patient represented for test purposes by a twenty centimetre water phantom. R.R.O. 1990, Reg. 543, s. 17 (4); O. Reg. 87/22, s. 7 (2).
(5) The exposure rate limit of a fluoroscopic x-ray machine that uses a cesium iodide input phosphor or a phosphor of similar efficiency calculated where the centre of the useful beam enters the patient at the shortest target-to-skin distance specified for the machine shall not exceed,
(a) a maximum exposure rate of ten roentgens per minute; and
(b) an entrance exposure rate of 2.5 roentgens per minute for an average patient represented for test purposes by a twenty centimetre water phantom. R.R.O. 1990, Reg. 543, s. 17 (5).
(6) Clauses (4) (a) and (5) (a) do not apply when the high-level control of the x-ray machine is activated. R.R.O. 1990, Reg. 543, s. 17 (6).
(7) A fluoroscopic x-ray machine that is equipped with an optional high-level control that allows higher exposure rates at the image receptor than the rates set out in subsections (4) and (5) shall be so constructed that,
(a) the high-level control is activated by its own control separate from any other control; and
(b) there is a continuous signal to the x-ray operator to indicate that the high-level control is being activated. R.R.O. 1990, Reg. 543, s. 17 (7).
(8) A photofluorographic x-ray machine shall only be used when the primary image is enhanced by electronic image intensification. R.R.O. 1990, Reg. 543, s. 17 (8).
Read the X-RAY SAFETY CODE on e-Laws ↗
Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.