X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543

Section 16

Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 16 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026

The text

Every general-purpose radiographic x-ray machine that is used with only one size of image receptor at a fixed target-to-image-receptor distance shall be equipped with devices to ensure that, (a) the centre of the x-ray field is aligned with the centre of the image receptor to within 2 per cent of the target-to-image-receptor distance; and (b) the x-ray field in the plane of the image receptor does not extend beyond any edge of the image receptor. R.R.O. 1990, Reg. 543, s. 16.
Read the X-RAY SAFETY CODE on e-Laws ↗

Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.