X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543
Section 18
Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 18 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026
The text
(1) Every x-ray machine that is used to conduct mammographic x-ray examinations shall be equipped with,
(a) an x-ray beam limiting device that limits the useful beam so that at any target-to-image-receptor distance specified for the machine the x-ray field in the plane of the image receptor,
(i) does not exceed the edge of the image receptor next to the chest wall by more than 2 per cent of the target-to-image-receptor distance, and
(ii) except for the edge of an image receptor referred to in subclause (i), does not extend beyond any other edge of the image receptor;
(b) an image receptor supporting device that is shielded sufficiently to ensure that for each activation of the x-ray tube the radiation exposure does not exceed 0.1 milliroentgens where,
(i) the machine is operated,
(A) in the mammographic mode,
(B) at the maximum rated x-ray tube potential,
(C) at the maximum rated tube current-exposure product for that tube potential, and
(D) at the minimum target-to-receptor distance attainable, and
(ii) the radiation exposure is averaged over a detection area of 100 square centimetres, with no linear dimension greater than twenty centimetres and centred at five centimetres from an accessible surface beyond the plane of the support device; and
(c) a device that will compress the breast of the patient being x-rayed. R.R.O. 1990, Reg. 543, s. 18 (1); O. Reg. 87/22, s. 8 (1).
(2) A removable fixed-operative beam limiting device that is installed on an x-ray machine that is constructed or adapted to perform mammographic examinations shall bear on its external surface clearly visible permanent markings that state,
(a) the image receptor size; and
(b) the target-to-image-receptor distance for which the beam limiting device is designed. R.R.O. 1990, Reg. 543, s. 18 (2).
(3) Every mammographic x-ray machine shall be so constructed that the accuracy of kilovoltage calibration for the machine is + 1 kilovolts for kilovoltage up to thirty-five and + 4 per cent for kilovoltage above thirty-five. R.R.O. 1990, Reg. 543, s. 18 (3).
(4) Every x-ray machine that is constructed or adapted to perform mammographic examinations shall be so constructed or adapted that,
(a) for any selected combination of kilovoltage, current and time, the estimated coefficient of variation of any ten consecutive radiation measurements taken at the same distance within a time period of one hour is not greater than 0.08;
(b) where the timer is non-mechanical, it is accurate to within 1/30 second (two cycles) or 10 per cent of the set value, whichever is greater; and
(c) where the timer is mechanical, it is accurate to within 1/20 second or 15 per cent, whichever is greater. R.R.O. 1990, Reg. 543, s. 18 (4); O. Reg. 87/22, s. 8 (2).
Read the X-RAY SAFETY CODE on e-Laws ↗
Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.