X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543

Section 15

Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 15 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026

The text

(1) Every general-purpose radiographic x-ray machine and every mobile radiographic x-ray machine shall be equipped with an x-ray beam limiting device that, (a) provides for stepless adjustment of the size of the x-ray field; (b) provides for a minimum field size that does not exceed five centimetres by five centimetres at a target-to-image-receptor distance of 100 centimetres; and (c) ensures that at each position, the x-ray field is aligned with the image receptor in such a manner that the x-ray field is always confined within the boundaries of the image receptor. (2) An x-ray beam limiting device referred to in subsection (1) shall, (a) be equipped with an adjustable light beam diaphragm or other device that defines visually the outline of the x-ray field when the axis of the x-ray beam is perpendicular to the plane of the image; or (b) allow the operator to adjust the dimensions of the x-ray field at the image receptor to a size smaller than the dimensions of the image receptor. (3) An adjustable light beam diaphragm or other device that defines visually the outline of the x-ray field shall be so constructed that, (a) misalignment of the visually defined field with respect to the x-ray field along either the length or width of the x-ray field does not exceed 2 per cent of the target-to-image-receptor distance; and (b) the size of the x-ray field in the plane of the image receptor is indicated at selected distances that are accurate to within 3 per cent of the target-to-image-receptor distance. R.R.O. 1990, Reg. 543, s. 15.
Read the X-RAY SAFETY CODE on e-Laws ↗

Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.