HEALTH CARE AND RESIDENTIAL FACILITIES, O. Reg. 67/93

Section 60

Cite as HEALTH CARE AND RESIDENTIAL FACILITIES, O. Reg. 67/93, s. 60 · current to November 11, 2026 · consolidated from June 9, 2025 · read from e-Laws on November 16, 2026

The text

(1) An autoclave or sterilization machine shall, (a) if the sterilant used in it is ethylene oxide or another hazardous chemical, be vented to the outdoors; (b) be operated by a worker qualified by training and experience to do so; (c) have its operating and emergency instructions posted adjacent to it; (d) be maintained on a regular basis and be inspected at least once every three months; (e) be tested when first put into service and annually thereafter by a person qualified by training and experience to do so and a record of the test shall be kept; and (f) have a pressure relief valve set at a pressure not exceeding that for which it was designed and intended. (2) After its operation or use, an autoclave or sterilization machine shall not be opened until its pressure has returned to atmospheric pressure. O. Reg. 67/93, s. 60.
Read the HEALTH CARE AND RESIDENTIAL FACILITIES on e-Laws ↗

Made under Occupational Health and Safety Act, R.S.O. 1990, c. O.1.