(1) A centrifuge shall be maintained and operated in accordance with the recommendations and instructions of its manufacturer. (2) A centrifuge shall be designed or equipped with a device to prevent it from being operated at a speed in excess of that for which it was designed and intended. (3) The load in a centrifuge shall be balanced to minimize vibration during its operation. (4) If a flammable or infectious material is being centrifuged, a legible sign warning of the hazard shall be posted in the area where the centrifuge is being operated. (5) If a bench-model centrifuge is being used to centrifuge a flammable or infectious material, it shall be operated in a biological safety cabinet or be otherwise appropriately contained, unless sealed safety heads or sealed centrifugal caps are used. (6) Cushions shall be used in centrifuging material that is in glass containers. O. Reg. 67/93, s. 59.