X-RAY SAFETY, R.R.O. 1990, Reg. 861
Section 18
Cite as X-RAY SAFETY, R.R.O. 1990, Reg. 861, s. 18 · current to November 11, 2026 · consolidated from July 1, 2021 · read from e-Laws on November 16, 2026
The text
In addition to any other requirements of this Regulation, where an employer is in possession of an X-ray source that consists of analytic X-ray equipment to which section 17 does not apply and that is primarily used to determine the structure or composition of a sample of a material, the employer shall comply with the following requirements:
1. The control panel shall have an indicator, in close proximity to the X-ray “ON/OFF” switch, that clearly indicates when X-rays are being produced in the X-ray tube.
2. A warning light shall be mounted near each X-ray tube in such a way as to be clearly visible from any direction from which the tube can be approached, that indicates when X-rays are being produced.
3. The condition of each shutter, open or closed, shall be clearly indicated at or near the X-ray tube.
4. Each port shall be designed in such a way that the X-ray beam can emerge only when a camera or other recording device is in its proper position, wherever practicable.
5. At least one of the warning or safety devices mentioned in paragraphs 1 to 4 shall be of failsafe design.
6. A guard or interlock which prevents entry of any part of the body into the primary beam path shall be used, wherever practicable.
7. A shield shall be provided to absorb the primary beam at the nearest practicable position beyond the point of intersection of the beam and the sample that it is intended to irradiate.
8. All unused ports shall be secured in such a way as to prevent inadvertent opening.