X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543

Section 3

Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 3 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026

The text

(1) Every installation of an x-ray machine shall be shielded with a primary protective barrier and a secondary protective barrier so that, (a) no x-ray worker receives a whole-body-dose-equivalent of more than 1 millisievert (100 millirem) per week; and (b) no person, other than the patient undergoing an application of therapeutic or diagnostic x-rays, who is not an x-ray worker, receives a whole-body-dose-equivalent of more than 0.1 millisievert (10 millirem) per week. R.R.O. 1990, Reg. 543, s. 3 (1). (2) The barriers referred to in subsection (1) shall comply with the shielding standards contained in one of the following: 1. Appendix II (Shielding guides for diagnostic x-ray installations) of Safety Code-20A—X-Ray Equipment in Medical Diagnosis Part A: Recommended Safety Procedures for Installation and Use, revised in 1999 and published by Health Canada. 2. NCRP Report No. 147—Structural Shielding Design for Medical X-Ray Imaging Facilities, revised on March 18, 2005 and published by the U.S. National Council on Radiation Protection and Measurements. O. Reg. 87/22, s. 2 (1). (3) Where lead shielding is used as a barrier, it shall be mounted in such a manner as to avoid sagging or damage to the lead shielding. R.R.O. 1990, Reg. 543, s. 3 (3). (4) Joints between different kinds of barrier material shall be constructed so that the overall attenuation of the barrier is not impaired. R.R.O. 1990, Reg. 543, s. 3 (4). (5) Windows, doors or other openings in a barrier shall be so constructed that they comply with one of the shielding standards referred to in subsection (2) that apply to barriers. O. Reg. 87/22, s. 2 (2). (6) All doors leading directly into an x-ray room shall be fitted with self-closing devices and, where the doors are accessible to the public, shall have prominently displayed on them warning signs sufficient to alert persons to the presence of the x-ray equipment. R.R.O. 1990, Reg. 543, s. 3 (6). (7) Unprocessed film shall be protected from x-rays being generated by x-ray machines in the facility so that during its storage the increase in optical density caused by unintentional irradiation is less than 0.02 density units. R.R.O. 1990, Reg. 543, s. 3 (7).
Read the X-RAY SAFETY CODE on e-Laws ↗

Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.