Occupational Health and Safety Regulation, B.C. Reg. 296/97
Section 30.12
Biological safety cabinets
Cite as Occupational Health and Safety Regulation, B.C. Reg. 296/97, s. 30.12 · last amended September 1, 2027 · read from BC Laws on November 17, 2026
The text
(1) The limitations of a biological safety cabinet must be clearly posted on the unit and followed by workers.
(2) Biological safety cabinets must be certified by a qualified person at least annually and before use after
(a) initial installation,
(b) change of the HEPA (high efficiency particulate air) filter,
(c) moving of the unit, and
(d) any repair or maintenance that could affect the seal of the HEPA filter.
(3) Certification procedures used for compliance with subsection (2) must meet the requirements of the National Sanitation Foundation (NSF) Standard 49-2002, Class II (Laminar Flow) Biohazard Cabinetry, and a record of the results must be maintained.
(4) Recirculation of exhaust air into a workspace from a biological safety cabinet is not permitted where volatile toxic materials or flammable liquids or gases are used in the cabinet, or where volatile radioactive materials are used in amounts that exceed the exemption quantity specified by the Canadian Nuclear Safety Commission.
(5) Repealed. [B.C. Reg. 319/2007, App. B, s. 6 (c).]
(6) Biological safety cabinets used for handling a biological agent must be operated and ventilated in accordance with the Laboratory Biosafety Guidelines 3rd edition, 2004, issued by the Public Health Agency of Canada.
Read the Occupational Health and Safety Regulation on BC Laws ↗
Made under Workers Compensation Act, RSBC 2019, c. 1.