S 1622 — FDA Device Accountability Act of 2016
Last action — Placed on Senate Legislative Calendar under General Orders. Calendar No. 426.
Write to your legislator about it: a free email template →
No card. One bill, free forever — alerts when it moves.
-
✓Introduced
-
✓In Committee
-
Died in session
This bill died with 114th Congress. It reached “In Committee” and never advanced before the session ended, so it can no longer move — a new version would have to be reintroduced in the current session.
This bill is no longer active — its legislative session has ended, so there is no live prognosis. It would have to be reintroduced in the current session to move again.
Summary
FDA Device Accountability Act of 2016 This bill amends the Federal Food, Drug, and Cosmetic Act to require the Food and Drug Administration (FDA) to ensure that employees who review premarket submissions of medical devices receive training on least burdensome requirements. (Currently, the FDA is required to consider the least burdensome appropriate means for a device sponsor to demonstrate the effectiveness of a device or its substantial equivalence to an approved device.) The FDA must periodically assess the implementation of those requirements. The ombudsman for any applicable unit of the FDA must conduct an audit of the training on least burdensome requirements. The FDA must consider: (1) the least burdensome appropriate means necessary to demonstrate device safety and effectiveness when requesting additional information from a device sponsor to support a premarket approval application, and (2) whether the least burdensome means would be reliance on postmarket information. The documentation of rationale for an applicable significant decision must include an explanation of how the least burdensome requirements were considered and applied. The Institutional Review Board responsible for reviewing the plan for the clinical testing of a medical device no longer needs to be local to the facilities where the testing will be conducted. The Department of Health and Human Services must revise its guidance entitled "Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices."
Bill Text
What changed in the latest version
202 added · 64 removedPlain-language change summary
The bill has been amended to remove the sections that established a short title and specific provisions regarding training, oversight, and ombudsman audits related to the evaluation of medical devices. The amendments streamline the content by eliminating detailed requirements about audits and training, focusing instead on essential principles without additional procedural steps. This change may simplify the process for evaluating medical devices by reducing the emphasis on compliance and auditing practices.
1622 IntroducedReported in Senate (IS)](RS)] 114th<DOC> CONGRESSCalendar 1stNo. Session S.
426 114th CONGRESS 2d Session S.
Burr (for himselfhimself, and Mr.
Franken)Franken, introducedMr. the following bill;
whichKirk, wasMr. read twice and referred to the Committee on Health, Education, Labor, and Pensions _______________________________________________________________________ A BILL To amend the Federal Food, Drug, and Cosmetic Act with respect to devices.
BeEnzi, it enacted by the Senate and HouseMr. of Representatives of the United States of America in Congress assembled, SECTION 1.
Alexander) introduced the following bill;
which was read twice and referred to the Committee on Health, Education, Labor, and Pensions April 18, 2016 Reported by Mr.
Alexander, with an amendment [Strike out all after the enacting clause and insert the part printed in italic] _______________________________________________________________________ A BILL To amend the Federal Food, Drug, and Cosmetic Act with respect to devices.
Be it enacted by the Senate and House of Representatives of the United States of America in Congress assembled, <DELETED>SECTION 1.
SHORT TITLE.</DELETED> <DELETED> This Act may be cited as the ``FDA Device Accountability Act of 2015''.</DELETED> <DELETED>SEC.
2.
ENSURING LEAST BURDENSOME MEANS OF EVALUATING DEVICES.</DELETED> <DELETED> (a) Training and Oversight of Least Burdensome Requirements.--Section 513 of the Federal Food, Drug, and Cosmetic Act (21 U.S.C.
360c) is amended by adding at the end the following:</DELETED> <DELETED> ``(j) Training and Oversight of Least Burdensome Requirements.--</DELETED> <DELETED> ``(1) Training and assessment.--The Secretary shall--</DELETED> <DELETED> ``(A) ensure that each employee of the Food and Drug Administration who is involved in the review of premarket submissions, including supervisors, receives training regarding the meaning and implementation of the least burdensome requirements under subsections (a)(3)(D) and (i)(1)(D) and section 515(c)(5);
and</DELETED> <DELETED> ``(B) periodically assess the implementation of the least burdensome requirements, including the employee training under subparagraph (A) to ensure that the least burdensome requirements are fully and consistently applied.</DELETED> <DELETED> ``(2) Ombudsman audit.--Not later than 180 calendar days after the date of enactment of the FDA Device Accountability Act of 2015, the ombudsman for any organizational unit of the Food and Drug Administration responsible for the premarket review of devices shall-- </DELETED> <DELETED> ``(A) conduct an audit of the training described in paragraph (1)(A);</DELETED> <DELETED> ``(B) include in such audit interviews of persons who are representatives of the device industry regarding their experience in the device premarket review process, including with respect to the application of least burdensome concepts to premarket review and the application of postmarket requirements to facilitate premarket decisionmaking;</DELETED> <DELETED> ``(C) include in such audit an assessment of the measurement tools the Secretary uses to assess the implementation of the least burdensome requirements, including the effectiveness of such tools and the effectiveness of the implementation of the least burdensome requirements;
and</DELETED> <DELETED> ``(D) within 30 calendar days of completion of the audit, make such audit available-- </DELETED> <DELETED> ``(i) to the Committee on Health, Education, Labor, and Pensions of the Senate and the Committee on Energy and Commerce of the House of Representatives;
and</DELETED> <DELETED> ``(ii) on the Internet website of the Food and Drug Administration.''.</DELETED> <DELETED> (b) Premarket Applications.--Section 515(c) of the Federal Food, Drug, and Cosmetic Act (21 U.S.C.
360e(c)) is amended by adding at the end the following:</DELETED> <DELETED> ``(5)(A) In requesting additional information with respect to an application under this section, the Secretary shall consider the least burdensome appropriate means necessary to demonstrate a reasonable assurance of device safety and effectiveness.</DELETED> <DELETED> ``(B) For purposes of subparagraph (A) the term `necessary' means the minimum required information that would support a determination by the Secretary that an application provides a reasonable assurance of the safety and effectiveness of the device.</DELETED> <DELETED> ``(C) Nothing in this paragraph alters the standards for premarket approval of a device.</DELETED> <DELETED> ``(D) For purposes of this paragraph, the Secretary shall consider whether the least burdensome means of demonstrating a reasonable assurance of device safety and effectiveness would be achieved through reliance on postmarket information.''.</DELETED> <DELETED> (c) Rationale for Significant Decisions Regarding Devices.--Section 517A(a) of the Federal Food, Drug, and Cosmetic Act (21 U.S.C.
360g-1(a)) is amended by adding at the end the following:</DELETED> <DELETED> ``(3) Application of least burdensome requirements.--The substantive summary required under this subsection shall include an explanation of how the least burdensome requirements were considered and applied consistent with section 513(i)(1)(D) and section 513(a)(3)(D) and section 515(c)(5), as applicable.''.</DELETED> <DELETED>SEC.
3.
PERMITTING NON-LOCAL INSTITUTIONAL REVIEW BOARDS.</DELETED> <DELETED> (a) In General.--Section 520 of the Federal Food, Drug, and Cosmetic Act (21 U.S.C.
360(j)) is amended--</DELETED> <DELETED> (1) in subsection (g)(3)--</DELETED> <DELETED> (A) by striking ``local'' each place it appears;
and</DELETED> <DELETED> (B) in subparagraph (A)(i), by striking ``which has been'';
and</DELETED> <DELETED> (2) in subsection (m)(4)--</DELETED> <DELETED> (A) by striking ``local'' each place it appears;
and</DELETED> <DELETED> (B) by amending subparagraph (A) to read as follows:</DELETED> <DELETED> ``(A) in facilities in which clinical testing of devices is supervised by an institutional review committee established in accordance with the regulations of the Secretary;
and''.</DELETED> <DELETED> (b) Regulations.--Not later than 1 year after the date of the enactment of this Act, the Secretary of Health and Human Services shall revise or issue such regulations or guidance as may be necessary to carry out the amendments made by subsection (a).</DELETED> <DELETED>SEC.
4.
CLARIFYING CLIA WAIVER STUDY DESIGN GUIDANCE FOR IN VITRO DIAGNOSTICS.</DELETED> <DELETED> (a) Draft Revised Guidance.--Not later than 1 year after the date of the enactment of this Act, the Secretary of Health and Human Services shall publish a draft guidance that--</DELETED> <DELETED> (1) revises section ``V.
Demonstrating Insignificant Risk of an Erroneous Result'' - ``Accuracy'' of the guidance entitled ``Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices'' and dated January 30, 2008;
and</DELETED> <DELETED> (2) includes guidance on the appropriate use of comparable performance between a waived user and a moderately complex laboratory user to demonstrate accuracy.</DELETED> <DELETED> (b) Final Revised Guidance.--The Secretary of Health and Human Services shall finalize the draft guidance published under subsection (a) not later than 1 year after the comment period for such draft guidance closes.</DELETED> SECTION 1.
This Act may be cited as the ``FDA Device Accountability Act of 2015''.2016''.
USE OF NONLOCAL INSTITUTIONAL REVIEW BOARDS FOR REVIEW OF INVESTIGATIONAL DEVICE EXEMPTIONS AND HUMAN DEVICE EXEMPTIONS.
Section 520 of the Federal Food, Drug, and Cosmetic Act (21 U.S.C.
360j) is amended-- (1) in subsection (g)(3)-- (A) in subparagraph (A)(i)-- (i) by striking ``local'';
and (ii) by striking ``which has been'';
and (B) in subparagraph (B), by striking ``a local institutional'' and inserting ``an institutional'';
and (2) in subsection (m)(4)-- (A) by striking subparagraph (A) and inserting the following new subparagraph:
``(A) in facilities in which clinical testing of devices is supervised by an institutional review committee established in accordance with the regulations of the Secretary, and'';
(B) in subparagraph (B), by striking ``a local institutional'' and inserting ``an institutional'';
and (C) in the matter following subparagraph (B), by striking ``local''.
Show all 77 changed lines (37 more)
SEC.
3.
CLIA WAIVER STUDY DESIGN GUIDANCE FOR IN VITRO DIAGNOSTICS.
(a) Draft Revised Guidance.--Not later than 1 year after the date of the enactment of this Act, the Secretary of Health and Human Services, acting through the Commissioner of Food and Drugs, shall publish a draft guidance that-- (1) revises section ``V.
Demonstrating Insignificant Risk of an Erroneous Result'' - ``Accuracy'' of the guidance entitled ``Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices'' and dated January 30, 2008;
and (2) includes the appropriate use of comparable performance between a waived user and a moderately complex laboratory user to demonstrate accuracy.
(b) Final Revised Guidance.--The Secretary of Health and Human Services, acting through the Commissioner of Food and Drugs, shall finalize the draft guidance published under subsection (a) not later than 1 year after the comment period for such draft guidance closes.
SEC.
4.
``(2) Ombudsman audit.--Not later than 18018 calendarmonths days after the date of enactment of the FDA Device Accountability Act of 2015,2016, the ombudsman for any organizational unit of the Food and Drug Administration responsible for the premarket review of devices shall-- ``(A) conduct an audit of the training described in paragraph (1)(A);(1)(A), including the effectiveness of such training in implementing the least burdensome requirements;
``(B) include in such audit interviews of persons who are representatives of the device industry regarding their experience in the device premarket review process, including with respect to the application of least burdensome concepts to premarket review and the application of postmarket requirements to facilitate premarket decisionmaking;
``(C) include in such audit ana assessmentlist of the measurement tools the Secretary uses to assess the implementation of the least burdensome requirements, including thethose effectivenessreferenced ofin suchparagraph tools(1)(B) and section 517A(a)(3), and may also provide feedback on the effectiveness of such tools in the implementation of the least burdensome requirements;
and ``(D) withinsummarize 30the calendarfindings days of completion of the audit, make such audit available--in ``(i)a tofinal theaudit Committeereport; on Health, Education, Labor, and Pensions of the Senate and the Committee on Energy and Commerce of the House of Representatives;
and ``(E) within 30 calendar days of completion of such final audit report, make such final audit report available-- ``(i) to the Committee on Health, Education, Labor, and Pensions of the Senate and the Committee on Energy and Commerce of the House of Representatives;
``(B) For purposes of subparagraph (A)(A), the term `necessary' means the minimum required information that would support a determination by the Secretary that an application provides a reasonable assurance of the safety and effectiveness of the device.
``(C) NothingFor inpurposes of this paragraphparagraph, altersthe Secretary shall consider the standardsrole forof premarketpostmarket approvalinformation in determining the least burdensome means of demonstrating a device.reasonable assurance of device safety and effectiveness.
``(D) ForNothing purposesin of this paragraph,paragraph themay Secretaryalter shall consider whether the leaststandards burdensomefor meanspremarket approval of demonstrating a reasonabledevice.''. assurance of device safety and effectiveness would be achieved through reliance on postmarket information.''.
``(3) Application of least burdensome requirements.--The substantive summary required under this subsection shall include ana explanationbrief ofstatement regarding how the least burdensome requirements were considered and applied consistent with section 513(i)(1)(D)513(i)(1)(D), and section 513(a)(3)(D)513(a)(3)(D), and section 515(c)(5), as applicable.''.
SEC.Calendar No.
3.426 114th CONGRESS 2d Session S.
PERMITTING1622 NON-LOCAL_______________________________________________________________________ INSTITUTIONALA REVIEWBILL BOARDS.To amend the Federal Food, Drug, and Cosmetic Act with respect to devices.
(a)_______________________________________________________________________ InApril General.--Section18, 5202016 ofReported thewith Federalan Food,amendment Drug, and Cosmetic Act (21 U.S.C.
360(j)) is amended-- (1) in subsection (g)(3)-- (A) by striking ``local'' each place it appears;
and (B) in subparagraph (A)(i), by striking ``which has been'';
and (2) in subsection (m)(4)-- (A) by striking ``local'' each place it appears;
and (B) by amending subparagraph (A) to read as follows:
``(A) in facilities in which clinical testing of devices is supervised by an institutional review committee established in accordance with the regulations of the Secretary;
and''.
(b) Regulations.--Not later than 1 year after the date of the enactment of this Act, the Secretary of Health and Human Services shall revise or issue such regulations or guidance as may be necessary to carry out the amendments made by subsection (a).
SEC.
4.
CLARIFYING CLIA WAIVER STUDY DESIGN GUIDANCE FOR IN VITRO DIAGNOSTICS.
(a) Draft Revised Guidance.--Not later than 1 year after the date of the enactment of this Act, the Secretary of Health and Human Services shall publish a draft guidance that-- (1) revises section ``V.
Demonstrating Insignificant Risk of an Erroneous Result'' - ``Accuracy'' of the guidance entitled ``Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices'' and dated January 30, 2008;
and (2) includes guidance on the appropriate use of comparable performance between a waived user and a moderately complex laboratory user to demonstrate accuracy.
(b) Final Revised Guidance.--The Secretary of Health and Human Services shall finalize the draft guidance published under subsection (a) not later than 1 year after the comment period for such draft guidance closes.
<all>
Show all 77 changed rows (37 more)
View plain text versions (2)
- Reported Reported to Senate Current html April 18, 2016
- Introduced Introduced in Senate html June 18, 2015
Sections of law this bill changes
5 sections in 1 lawFrom the bill's own amending language, matched to the official United States Code (current through Public Law 119-142).
Federal food, drug, and cosmetic act · 21 U.S.C. ch. 9 5 sections
-
§ 360 Registration of producers of drugs or devices Amends
21 U.S.C. § 360 — at 360(j)
The law today
(a) Definitions As used in this section— (1) the term “manufacture, preparation, propagation, compounding, or processing” shall include repackaging or otherwise changing the container, wrapper, or labeling of any drug package or device package in furtherance of the distribution of the drug or device from the original place of manufacture to the person who makes final delivery or sale to the ultimate consumer or user; and (2) the term “name” shall include in the case of a partnership the name of each partner and, in the case of a corporation, the name of each corporate officer and director, and the State of incorporation. (b) Annual registration (1) During the period beginning on October 1 and ending on December 31 of each year, every person who owns or operates any establishment in any State engaged in the manufacture, preparation, propagation, compounding, or processing of a drug or drugs shall register with the Secretary the name of such person, places of business of such person, all such establishments, the unique facility identifier of each such establishment, and a point of contact e-mail address. (2) During the period beginning on October 1 and ending on December 31 of each year, every person who owns or operates any establishment in any State engaged in the manufacture, preparation, propagation, compounding, or processing of a device or devices shall register with the Secretary his name, places of business, and all such establishments. (3) The Secretary shall specify the unique facility identifier system that shall be used by registrants under paragraph (1). The requirement to include a unique facility identifier in a registration under paragraph (1) shall not apply until the date that the identifier system is specified by the Secretary under the preceding sentence. (c) New producers Every person upon first engaging in the manufacture, preparation, propagation, compounding, or processing of a drug or drugs or a device or devices in any establishment which he owns or operates in any State shall immediately register with the Secretary— (1) with respect to drugs, the information described under subsection (b)(1); and (2) with respect to devices, the information described under subsection (b)(2)..1 (d) Additional establishments Every person duly registered in accordance with the foregoing subsections of this section shall immediately register with the Secretary any additional establishment which he owns or operates in any State and in which he begins the manufacture, preparation, propagation, compounding, or processing of a drug or drugs or a device or devices. (e) Registration number; uniform system for identification of devices intended for human use The Secretary may assign a registration number to any person or any establishment registered in accordance with this section. The Secretary may also assign a listing number to each drug or class of drugs listed under subsection (j). Any number assigned pursuant to the preceding sentence shall be the same as that assigned pursuant to the National Drug Code. The Secretary may by regulation prescribe a uniform system for the identification of devices intended for human use and may require that persons who are required to list such devices pursuant to subsection (j) shall list such devices in accordance with such system. (f) Availability of registrations for inspection The Secretary shall make available for inspection, to any person so requesting, any registration filed pursuant to this section; except that any list submitted pursuant to paragraph (3) of subsection (j) and the information accompanying any list or notice filed under paragraph (1) or (2) of that subsection shall be exempt from such inspection unless the Secretary finds that such an exemption would be inconsistent with protection of the public health. (g) Exclusions from application of section The foregoing subsections of this section shall not apply to— (1) pharmacies which maintain establishments in conformance...79 other bills also change this section · Follow it → Official text at uscode.house.gov ↗
-
§ 360c Classification of devices intended for human use Amends
21 U.S.C. § 360c
The law today
(a) Classes of devices (1) There are established the following classes of devices intended for human use: (A) Class I, General Controls.— (i) A device for which the controls authorized by or under section 351, 352, 360, 360f, 360h, 360i, or 360j of this title or any combination of such sections are sufficient to provide reasonable assurance of the safety and effectiveness of the device. (ii) A device for which insufficient information exists to determine that the controls referred to in clause (i) are sufficient to provide reasonable assurance of the safety and effectiveness of the device or to establish special controls to provide such assurance, but because it— (I) is not purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, and (II) does not present a potential unreasonable risk of illness or injury, is to be regulated by the controls referred to in clause (i). (B) Class II, Special Controls.— A device which cannot be classified as a class I device because the general controls by themselves are insufficient to provide reasonable assurance of the safety and effectiveness of the device, and for which there is sufficient information to establish special controls to provide such assurance, including the promulgation of performance standards, postmarket surveillance, patient registries, development and dissemination of guidelines (including guidelines for the submission of clinical data in premarket notification submissions in accordance with section 360(k) of this title), recommendations, and other appropriate actions as the Secretary deems necessary to provide such assurance. For a device that is purported or represented to be for a use in supporting or sustaining human life, the Secretary shall examine and identify the special controls, if any, that are necessary to provide adequate assurance of safety and effectiveness and describe how such controls provide such assurance. (C) Class III, Premarket Approval.— A device which because— (i) it (I) cannot be classified as a class I device because insufficient information exists to determine that the application of general controls are sufficient to provide reasonable assurance of the safety and effectiveness of the device, and (II) cannot be classified as a class II device because insufficient information exists to determine that the special controls described in subparagraph (B) would provide reasonable assurance of its safety and effectiveness, and (ii) (I) is purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, or (II) presents a potential unreasonable risk of illness or injury, is to be subject, in accordance with section 360e of this title, to premarket approval to provide reasonable assurance of its safety and effectiveness. If there is not sufficient information to establish a performance standard for a device to provide reasonable assurance of its safety and effectiveness, the Secretary may conduct such activities as may be necessary to develop or obtain such information. (2) For purposes of this section and sections 360d and 360e of this title, the safety and effectiveness of a device are to be determined— (A) with respect to the persons for whose use the device is represented or intended, (B) with respect to the conditions of use prescribed, recommended, or suggested in the labeling of the device, and (C) weighing any probable benefit to health from the use of the device against any probable risk of injury or illness from such use. (3) (A) Except as authorized by subparagraph (B), the effectiveness of a device is, for purposes of this section and sections 360d and 360e of this title, to be determined, in accordance with regulations promulgated by the Secretary, on the basis of well-controlled investigations, including 1 or more clinical in...25 other bills also change this section · Follow it → Official text at uscode.house.gov ↗
-
§ 360e Premarket approval Amends
21 U.S.C. § 360e — at 360e(c)
The law today
(a) General requirement A class III device— (1) which is subject to an order issued under subsection (b) (or a regulation promulgated under such subsection prior to July 9, 2012); or (2) which is a class III device because of section 360c(f) of this title, is required to have, unless exempt under section 360j(g) of this title, an approval under this section of an application for premarket approval or, as applicable, an approval under subsection (c)(2) of a report seeking premarket approval. (b) Order to require premarket approval (1) In the case of a class III device which— (A) was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976; or (B) is (i) of a type so introduced or delivered, and (ii) is substantially equivalent to another device within that type, the Secretary shall by administrative order following publication of a proposed order in the Federal Register, a meeting of a device classification panel described in section 360c(b) of this title, and consideration of comments from all affected stakeholders, including patients, payors, and providers, notwithstanding subchapter II of chapter 5 of title 5, require that such device have an approval under this section of an application for premarket approval. Authority to issue such administrative order shall not be delegated below the Director of the Center for Devices and Radiological Health, acting in consultation with the Commissioner. (2) A proposed order required under paragraph (1) shall contain— (A) the proposed order; (B) proposed findings with respect to the degree of risk of illness or injury designed to be eliminated or reduced by requiring the device to have an approved application for premarket approval and the benefit to the public from use of the device; (C) opportunity for the submission of comments on the proposed order and the proposed findings; and (D) opportunity to request a change in the classification of the device based on new information relevant to the classification of the device. (3) After the expiration of the period for comment on a proposed order and proposed findings published under paragraph (2), consideration of comments submitted on such proposed order and findings, and a meeting of a device classification panel described in section 360c(b) of this title, the Secretary shall (A) issue an administrative order under paragraph (1) and publish in the Federal Register findings on the matters referred to in paragraph (2)(B), or (B) publish a notice terminating the proceeding for the issuance of the administrative order together with the reasons for such termination. If a notice of termination is published, the Secretary shall (unless such notice is issued because the device is a banned device under section 360f of this title) initiate a proceeding under section 360c(e) of this title to reclassify the device subject to the proceeding terminated by such notice. (c) Application for premarket approval (1) Any person may file with the Secretary an application for premarket approval for a class III device. Such an application for a device shall contain— (A) full reports of all information, published or known to or which should reasonably be known to the applicant, concerning investigations which have been made to show whether or not such device is safe and effective; (B) a full statement of the components, ingredients, and properties and of the principle or principles of operation, of such device; (C) a full description of the methods used in, and the facilities and controls used for, the manufacture, processing, and, when relevant, packing and installation of, such device; (D) an identifying reference to any performance standard under section 360d of this title which would be applicable to any aspect of such device if it were a class II device, and either adequate information to show that such aspect of such device fully meets such performance standard or adequate information to justify an...21 other bills also change this section · Follow it → Official text at uscode.house.gov ↗
-
§ 360g-1 Agency documentation and review of significant decisions regarding devices Amends
21 U.S.C. § 360g-1 — at 360g-1(a)
The law today
(a) Documentation of rationale for significant decisions (1) In general The Secretary shall provide a substantive summary of the scientific and regulatory rationale for any significant decision of the Center for Devices and Radiological Health regarding submission or review of a report under section 360(k) of this title, a petition for classification under section 360c(f) of this title, an application under section 360e of this title, or an application for an exemption under section 360j(g) of this title, including documentation of significant controversies or differences of opinion and the resolution of such controversies or differences of opinion. (2) Provision of documentation Upon request, the Secretary shall furnish such substantive summary to the person who is seeking to submit, or who has submitted, such report or application. (3) Application of least burdensome requirements The substantive summary required under this subsection shall include a brief statement regarding how the least burdensome requirements were considered and applied consistent with section 360c(i)(1)(D) of this title, section 360c(a)(3)(D) of this title, and section 360e(c)(5) of this title, as applicable. (b) Review of significant decisions (1) Request for supervisory review of significant decision Any person may request a supervisory review of the significant decision described in subsection (a)(1). Such review may be conducted at the next supervisory level or higher above the individual who made the significant decision. (2) Submission of request A person requesting a supervisory review under paragraph (1) shall submit such request to the Secretary not later than 30 days after such decision and shall indicate in the request whether such person seeks an in-person meeting or a teleconference review. (3) Timeframe (A) In general Except as provided in subparagraph (B), the Secretary shall schedule an in-person or teleconference review, if so requested, not later than 30 days after such request is made. The Secretary shall issue a decision to the person requesting a review under this subsection not later than 45 days after the request is made under paragraph (1), or, in the case of a person who requests an in-person meeting or teleconference, 30 days after such meeting or teleconference. (B) Exception Subparagraph (A) shall not apply in cases that are referred to experts outside of the Food and Drug Administration.7 other bills also change this section · Follow it → Official text at uscode.house.gov ↗
-
§ 360j General provisions respecting control of devices intended for human use Amends
21 U.S.C. § 360j
The law today
(a) General rule Any requirement authorized by or under section 351, 352, 360, or 360i of this title applicable to a device intended for human use shall apply to such device until the applicability of the requirement to the device has been changed by action taken under section 360c, 360d, or 360e of this title or under subsection (g) of this section, and any requirement established by or under section 351, 352, 360, or 360i of this title which is inconsistent with a requirement imposed on such device under section 360d or 360e of this title or under subsection (g) of this section shall not apply to such device. (b) Custom devices (1) In general The requirements of sections 360d and 360e of this title shall not apply to a device that— (A) is created or modified in order to comply with the order of an individual physician or dentist (or any other specially qualified person designated under regulations promulgated by the Secretary after an opportunity for an oral hearing); (B) in order to comply with an order described in subparagraph (A), necessarily deviates from an otherwise applicable performance standard under section 360d of this title or requirement under section 360e of this title; (C) is not generally available in the United States in finished form through labeling or advertising by the manufacturer, importer, or distributor for commercial distribution; (D) is designed to treat a unique pathology or physiological condition that no other device is domestically available to treat; (E) (i) is intended to meet the special needs of such physician or dentist (or other specially qualified person so designated) in the course of the professional practice of such physician or dentist (or other specially qualified person so designated); or (ii) is intended for use by an individual patient named in such order of such physician or dentist (or other specially qualified person so designated); (F) is assembled from components or manufactured and finished on a case-by-case basis to accommodate the unique needs of individuals described in clause (i) or (ii) of subparagraph (E); and (G) may have common, standardized design characteristics, chemical and material compositions, and manufacturing processes as commercially distributed devices. (2) Limitations Paragraph (1) shall apply to a device only if— (A) such device is for the purpose of treating a sufficiently rare condition, such that conducting clinical investigations on such device would be impractical; (B) production of such device under paragraph (1) is limited to no more than 5 units per year of a particular device type, provided that such replication otherwise complies with this section; and (C) the manufacturer of such device notifies the Secretary on an annual basis, in a manner prescribed by the Secretary, of the manufacture of such device. (3) Guidance Not later than 2 years after July 9, 2012, the Secretary shall issue final guidance on replication of multiple devices described in paragraph (2)(B). (c) Trade secrets Any information reported to or otherwise obtained by the Secretary or his representative under section 360c, 360d, 360e, 360f, 360h, 360i, or 374 of this title or under subsection (f) or (g) of this section which is exempt from disclosure pursuant to subsection (a) of section 552 of title 5 by reason of subsection (b)(4) of such section shall be considered confidential and shall not be disclosed and may not be used by the Secretary as the basis for the reclassification of a device from class III to class II or class I or as the basis for the establishment or amendment of a performance standard under section 360d of this title for a device reclassified from class III to class II, except (1) in accordance with subsection (h), and (2) that such information may be disclosed to other officers or employees concerned with carrying out this chapter or when relevant in any proceeding under this chapter (other than section 360c or 360d of this title). (d) Notices and find...40 other bills also change this section · Follow it → Official text at uscode.house.gov ↗
Also cites: 21 U.S.C. § 360g
Action History
-
Placed on Senate Legislative Calendar under General Orders. Calendar No. 426.
-
Committee on Health, Education, Labor, and Pensions. Reported by Senator Alexander with an amendment in the nature of a substitute. Without written report.
-
Committee on Health, Education, Labor, and Pensions. Ordered to be reported with an amendment in the nature of a substitute favorably.
-
Read twice and referred to the Committee on Health, Education, Labor, and Pensions.
Sponsors
- Al Franken · Cosponsor
- Mark Steven Kirk · Cosponsor
- Michael B. Enzi · Cosponsor
- Lamar Alexander · Cosponsor
- Richard Burr · Primary
Sponsorship breakdown
Export CSV (upgrade) →1 sponsors · 4 co-sponsors · 543 not signed on
Sponsors (1)
Co-sponsors (4)
- Al Franken Democratic
- Mark Steven Kirk Republican
- Michael B. Enzi Republican
- Lamar Alexander Republican
Not signed on (543)
543 members have not signed on to this bill.
Show all 543 →"Not signed on" means a member has not sponsored or co-sponsored this bill — it does not imply opposition. Members flagged Voted No have a recorded No vote on this bill.
Frequently asked questions
- What does S 1622 do?
- FDA Device Accountability Act of 2016 This bill amends the Federal Food, Drug, and Cosmetic Act to require the Food and Drug Administration (FDA) to ensure that employees who review premarket submissions of medical devices receive training on least burdensome requirements. (Currently, the FDA is required to consider the least burdensome appropriate means for a device sponsor to demonstrate the effectiveness of a device or its substantial equivalence to an approved device.) The FDA must periodically assess the implementation of those requirements. The ombudsman for any applicable unit of the FDA must conduct an audit of the training on least burdensome requirements. The FDA must consider: (1) the least burdensome appropriate means necessary to demonstrate device safety and effectiveness when requesting additional information from a device sponsor to support a premarket approval application, and (2) whether the least burdensome means would be reliance on postmarket information. The documentation of rationale for an applicable significant decision must include an explanation of how the least burdensome requirements were considered and applied. The Institutional Review Board responsible for reviewing the plan for the clinical testing of a medical device no longer needs to be local to the facilities where the testing will be conducted. The Department of Health and Human Services must revise its guidance entitled "Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices."
- Who sponsors S 1622?
- S 1622 is sponsored by Al Franken (Democratic), Mark Steven Kirk (Republican), Michael B. Enzi (Republican), Lamar Alexander (Republican), and Richard Burr (R).
- What is the current status of S 1622?
- This bill died with 114th Congress. It reached “In Committee” and never advanced before the session ended, so it can no longer move — a new version would have to be reintroduced in the current session.
- Where can I track S 1622?
- Track S 1622 free on One Click Politics — get push/email alerts when it moves.
Make your voice heard on S 1622
Find the representatives who decide this bill and tell them where you stand — for yourself, or mobilize your whole list in one click with One Click Politics advocacy software.
Stay ahead of S 1622
Last checked for changes 16 days ago · updated continuously
One Click Politics tracks every bill in Congress and all 50 states.
Track this bill →