Health and Human Services Department Docket FDA-2026-N-6710 Final Rule

2026-12905: Medical Devices; General Hospital and Personal Use Devices; Classification of the Infant Supine Sleep System

Get alerted when the next rule lands

New rules open for comment on their own schedule. Pick how you want to hear about them — we email you when one appears.

Publishing to members? Embed a live rule tracker on your own site — the proposed rules in your area and the comment deadlines closing on them, updating themselves. Free, one line of HTML.

Create a free account & build one →

Summary

The Food and Drug Administration (FDA) is classifying the infant supine sleep system into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the infant supine sleep system. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

The agency’s own summary, as published.

The rule, in full

2,188 words as published, June 26, 2026. View the original →

Federal Register, Volume 91 Issue 122 (Friday, June 26, 2026) [Federal Register Volume 91, Number 122 (Friday, June 26, 2026)] [Rules and Regulations] [Pages 38506-38508] From the Federal Register Online via the Government Publishing Office [www.gpo.gov] [FR Doc No: 2026-12905] ----------------------------------------------------------------------- DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration 21 CFR Part 880 [Docket No. FDA-2026-N-6710] Medical Devices; General Hospital and Personal Use Devices; Classification of the Infant Supine Sleep System

Agency

Food and Drug Administration, HHS.

Action

Final amendment; final order. -----------------------------------------------------------------------

Summary

The Food and Drug Administration (FDA) is classifying the infant supine sleep system into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the infant supine sleep system. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

Dates

This order is effective June 26, 2026. The classification was applicable on March 30, 2023.

For Further Information Contact

Kathleen Everett, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 2522, Silver Spring, MD 20993-0002, 301- 796-6292, [email protected].

Supplementary Information

I. Background

Upon request, FDA (the Agency or we) has classified the infant supine sleep system into class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness of the device. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens by placing the device into a lower device class than the automatic class III assignment. The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified into, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (21 U.S.C. 360c(f)(1)). We refer to these devices as ``postamendments devices'' because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act). FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807). FDA may also classify a device through ``De Novo'' classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)). Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo classification process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification. Under the first procedure, the person submits a premarket notification (510(k)) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2). Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a [[Page 38507]] classification under section 513(f)(2) of the FD&C Act. Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device. We believe this De Novo classification will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens. When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining ``substantial equivalence''). Instead, sponsors can use the less burdensome 510(k) process, when necessary, to market their device.

II. De Novo Classification

On September 20, 2021, FDA received Happiest Baby, Inc.'s request for De Novo classification of the SNOO Smart Sleeper. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act. We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness of the device, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see section 513(a)(1)(B) of the FD&C Act). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device. Therefore, on March 30, 2023, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 880.5690.\1\ We have named the generic type of device ``infant supine sleep system,'' and it is identified as a device intended to facilitate a supine position during sleep for use in infants that are not yet able to roll over consistently. Infants placed in a supine sleep position are at lower risk of sudden infant death syndrome or sudden unexpected infant death. --------------------------------------------------------------------------- \1\ FDA notes that the ``ACTION'' caption for this final order is styled as ``Final amendment; final order,'' rather than ``Final order.'' Beginning in December 2019, this editorial change was made to indicate that the document ``amends'' the Code of Federal Regulations. The change was made in accordance with the Office of Federal Register's (OFR) interpretations of the Federal Register Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook. --------------------------------------------------------------------------- FDA has identified the risks to health associated with this type of device and the measures required to mitigate these risks in table 1. Table 1--Risks to Health and Mitigation Measures for Infant Supine Sleep System ------------------------------------------------------------------------ Identified risks to health Mitigation measures ------------------------------------------------------------------------ Increased risk of death, including from Clinical data; Postmarket inadequate securement or inadequate surveillance; Non-clinical positioning of the infant. performance testing; and labeling. Inappropriate securement leading to: Clinical data; Postmarket Injuries, contusions, or surveillance; and labeling. bruising Entrapment Respiratory compromise or suffocation Gastroesophageal reflux Plagiocephaly (``flat head syndrome'') Death Inappropriate or inadequate securement Non-clinical performance due to device degradation over time testing; and labeling. (wear and tear, laundering). Inappropriate use or inadequate Clinical data; Human factors securement due to use error and/or assessment; and labeling. improper fit. Injury due to unstable device (tipping, Non-clinical performance rocking, improper placement). testing; and labeling. Infection.............................. Labeling. Adverse tissue reaction (e.g., Biocompatibility evaluation; dermatitis). and labeling. ------------------------------------------------------------------------ FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness of the device. For a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this final order. Under the FD&C Act, submission of a premarket notification under section 510(k) is required to reasonably assure the safety and effectiveness of class II devices unless FDA determines that the device type should be exempt under section 510(m) of the FD&C Act. At this time FDA has not made this determination for infant supine sleep systems. This device is therefore subject to premarket notification requirements under section 510(k) of the FD&C Act.

III. Analysis of Environmental Impact

The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not normally have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

IV. Paperwork Reduction Act of 1995

This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations and guidance. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections of information in part 860, subpart D, regarding De Novo classification have been approved under OMB control number 0910-0844; the [[Page 38508]] collections of information in 21 CFR part 814, subparts A through E, regarding premarket approval have been approved under OMB control number 0910-0231; the collections of information in part 807, subpart E, regarding premarket notification submissions have been approved under OMB control number 0910-0120; the collections of information in 21 CFR part 820 regarding quality management system regulation have been approved under OMB control number 0910-0073; and the collections of information in 21 CFR part 801 regarding labeling have been approved under OMB control number 0910-0485. List of Subjects in 21 CFR Part 880 Medical devices. Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 880 is amended as follows: PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES 0

1. The authority citation for part 880 continues to read as follows:

Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371. 0

2. Add Sec. 880.5690 to subpart F to read as follows:

Sec. 880.5690 Infant supine sleep system. (a) Identification. An infant supine sleep system is a device intended to facilitate a supine position during sleep for use in infants that are not yet able to roll over consistently. Infants placed in a supine sleep position are at lower risk of sudden infant death syndrome (SIDS) or sudden unexpected infant death (SUID). (b) Classification. Class II (special controls). The special controls for this device are: (1) Premarket clinical information and, as determined by FDA, postmarket surveillance data acquired under anticipated conditions of use must be collected to fulfill the following: (i) Demonstrate that the device holds the infant on the back; (ii) Provide data on adverse events (including deaths and injuries) and malfunctions to demonstrate the device can be safely used in the intended use population; and (iii) Provide data to demonstrate that use of the device does not increase the rate of SIDS/SUID in the intended use population. (2) Human factors testing must demonstrate that the user can safely and correctly use the device. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be conducted: (i) Testing to ensure the mechanical and structural stability of the device and demonstrate that the device does not present a tipping hazard due to mechanical failures; and (ii) Material compatibility testing to demonstrate that the cleaning instructions provided by the manufacturer do not cause crazing, cracking, or deterioration of the device. (5) Labeling must include: (i) Unless clinical performance data demonstrates that it can be removed or modified, a prominent warning that the device has not been demonstrated to reduce the risk of SIDS/SUID. Such warning must appear prominently on all labeling; (ii) A summary of available clinical information with the device, including a discussion of adverse events; (iii) A warning that the device is only indicated for use with infants who cannot consistently roll over; (iv) Instructions to ensure proper fit; (v) Instructions for cleaning the device; and (vi) Information regarding safe sleep practices to ensure the safe use of the device, including: (A) Recommendations for safe sleep environments; and (B) The level of supervision necessary to monitor a sleeping infant. Grace R. Graham, Deputy Commissioner for Policy, Legislation, and International Affairs. [FR Doc. 2026-12905 Filed 6-25-26; 8:45 am] BILLING CODE 4164-01-P