CONSTRUCTION PROJECTS, O. Reg. 213/91
Section 357
Cite as CONSTRUCTION PROJECTS, O. Reg. 213/91, s. 357 · current to November 11, 2026 · consolidated from April 20, 2026 · read from e-Laws on November 16, 2026
The text
(1) A medical lock shall be not less than 1.8 metres high at its centre line. O. Reg. 213/91, s. 357 (1).
(2) A medical lock shall be divided into two pressure compartments. O. Reg. 213/91, s. 357 (2).
(3) Each compartment of a medical lock shall have air valves that are arranged so that the compartment can be pressurized and depressurized from inside and outside the lock. O. Reg. 213/91, s. 357 (3).
(4) An observation window shall be installed in each door and in the rear wall of a medical lock. O. Reg. 213/91, s. 357 (4).
(5) A medical lock shall be equipped with,
(a) a pressure release valve which will automatically blow-off at a pressure not greater than seventy kilopascals more than the operating pressure of the work chamber;
(b) a pressure gauge, a thermometer, a telephone, a cot, seating and a radiant heater; and
(c) a cot mattress, mattress cover and blankets all of which are made of material that is not readily flammable. O. Reg. 213/91, s. 357 (5).
(6) The pressure release valve shall be tested and calibrated before the medical lock is used. O. Reg. 213/91, s. 357 (6).
(7) A medical lock shall be maintained at a temperature of at least 18 degrees celsius, well-lit and well-ventilated and kept clean and sanitary. O. Reg. 213/91, s. 357 (7).