X-RAY SAFETY, R.R.O. 1990, Reg. 861
Section 11
Cite as X-RAY SAFETY, R.R.O. 1990, Reg. 861, s. 11 · current to November 11, 2026 · consolidated from July 1, 2021 · read from e-Laws on November 16, 2026
The text
The following measures and procedures shall be carried out in a workplace where an X-ray source is used:
1. X-ray warning signs or warning devices shall be posted or installed in conspicuous locations.
2. Every X-ray source capable of producing an air kerma rate greater than 5 micrograys per hour at any accessible point shall be labelled at its operating controls as a source of X-rays.
3. Where the air kerma in an area may exceed 100 micrograys in any one hour, access to the area shall be controlled by,
i. locks or interlocks if the X-ray source is one to which subsection 6 (1) applies or is described in subsection 6 (2), and
ii. barriers and X-ray warning signs if the X-ray source is portable or mobile and is being so used.
4. To ensure that the dose equivalent limits mentioned in section 10 are not exceeded,
i. structural or other shielding shall be installed as is necessary, and
ii. diaphragms, cones and adjustable collimators or other suitable devices shall be provided and used as are necessary to limit the dimensions of the useful X-ray beam.
Read the X-RAY SAFETY on e-Laws ↗
Made under Occupational Health and Safety Act, R.S.O. 1990, c. O.1.