X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543
Section 9
Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 9 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026
The text
(1) Every diagnostic x-ray machine shall bear either on the external surface of the main x-ray control panel or at the exposure switch location a warning sign that indicates that,
(a) unauthorized use is prohibited; and
(b) hazardous radiation is emitted when the x-ray machine is activated. R.R.O. 1990, Reg. 543, s. 9 (1).
(2) Every diagnostic x-ray machine shall be so constructed that,
(a) all controls, meters, lights or other indicators on the machine are readily recognizable and clearly identifiable as to function;
(b) the x-ray tube is securely fixed and correctly aligned with the tube housing;
(c) the x-ray tube housing maintains its required exposure position without significant drifting, tipping or vibration so as to affect the quality of the image;
(d) there are recognizable warning lights or other indicators that indicate,
(i) when the machine is energized and is ready to produce x-rays, and
(ii) when the x-rays are produced;
(e) where the machine has individual technique factors that are either fixed or can be selected manually by the x-ray operator, there are electrical meters, controls or other indicators to enable the x-ray operator to determine those selected technique factors before the patient is irradiated;
(f) where the x-ray machine is used in the radiographic mode and has automatically controlled exposure or anatomically related exposure selection or falling load, there is an electrical meter, control or other indicator that enables the x-ray operator to determine the kilovoltage before the patient is irradiated;
(g) where the x-ray machine is battery powered, there is a visual indicator that shows whether the battery is charged for proper operation;
(h) it is not possible to energize more than one x-ray tube at the same time; and
(i) where there are two x-ray tubes, there is a visible indication of which x-ray tube is selected and ready to be activated at the control panel. R.R.O. 1990, Reg. 543, s. 9 (2); O. Reg. 87/22, s. 6.
(3) Every diagnostic x-ray machine shall be provided with,
(a) an exposure switch, timer or other device that is controlled by the operator to initiate and terminate the irradiations; and
(b) filters that,
(i) are located in the exit port of the x-ray tube housing or beam limiting device or both,
(ii) intercept the entire useful beam, and
(iii) at a measured potential set out in Column 1 of Table 8 with a thickness of aluminum set out opposite thereto in Column 2 of Table 8, reduce the exposure at least by half. R.R.O. 1990, Reg. 543, s. 9 (3).
Read the X-RAY SAFETY CODE on e-Laws ↗
Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.