X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543
Section 20
Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 20 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026
The text
(1) Every dental x-ray machine and every chiropodic x-ray machine shall be so constructed that,
(a) for any selected combination of kilovoltage, current and time, the estimated coefficient of variation of any ten consecutive radiation measurements taken at the same distance within a time period of one hour is not greater than 0.08;
(b) when the x-ray machine is operating in the fixed milliamperage mode, the timer is, at each setting, accurate to within 1/30 second (two cycles) or 10 per cent of the set value; and
(c) for any selected setting of the peak x-ray tube potential, the actual peak kilovoltage corresponds to the selected value to within + 8 per cent.
(2) Clause (1) (b) does not apply to equipment used for panoramic dental examinations. R.R.O. 1990, Reg. 543, s. 20.
Read the X-RAY SAFETY CODE on e-Laws ↗
Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.