X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543
Section 11
Cite as X-RAY SAFETY CODE, R.R.O. 1990, Reg. 543, s. 11 · current to November 11, 2026 · consolidated from January 1, 2026 · read from e-Laws on November 16, 2026
The text
(1) Every diagnostic x-ray machine and every fluoroscopic x-ray machine shall, except where the x-ray machine is equipped with an automatic exposure control device, be so constructed that the timing device on the machine terminates an irradiation on completion of,
(a) a preset time interval;
(b) a preset product of current and time; or
(c) a preset number of pulses.
(2) Where an x-ray machine is equipped with an automatic exposure control device, the device shall terminate the exposure to the patient when a predetermined amount of radiation is detected.
(3) Every timing device on a diagnostic x-ray machine and fluoroscopic x-ray machine shall be so constructed that it,
(a) resets automatically to its original position or to ZERO on termination of an irradiation; and
(b) prevents an irradiation from occurring at the ZERO or OFF position. R.R.O. 1990, Reg. 543, s. 11.
Read the X-RAY SAFETY CODE on e-Laws ↗
Made under Healing Arts Radiation Protection Act, R.S.O. 1990, c. H.2.