Medical Devices Regulations, SOR/98-282

Section 65

Cite as Medical Devices Regulations, SOR/98-282, s. 65 · current to June 17, 2026 · regulation last amended June 17, 2026

Statute Watch

Get an alert when a bill in Parliament would change this section, and every time it moves.

The text

A manufacturer or importer of a medical device shall, within 30 days after completing a recall of the device that was not ordered by the Minister, provide the Minister with the following information, in writing: (a) the results of the recall; and (b) the action taken to prevent a recurrence of the problem.
Read the Medical Devices Regulations on the Justice Laws Website ↗

Made under Food and Drugs Act, R.S.C. 1985, c. F-27.