Medical Devices Regulations, SOR/98-282

Section 63.2

Cite as Medical Devices Regulations, SOR/98-282, s. 63.2 · current to June 17, 2026 · regulation last amended June 17, 2026

Statute Watch

Get an alert when a bill in Parliament would change this section, and every time it moves.

The text

A manufacturer or importer of a medical device who decides to recall the device without being ordered to do so by the Minister shall provide the Minister with the following information, in writing, within 24 hours after making the decision: (a) the name of the device; (b) the identifier of the device, including the identifier of any medical device that is part of a system, test kit, medical device group, medical device family or medical device group family; (c) in the case of a licensed device, the medical device licence number; (d) in the case of a device for which the manufacturer holds an authorization issued under section 68.12, the authorization number; (e) the name and address of (i) the manufacturer, (ii) the establishment where the device was manufactured, if different from that of the manufacturer, and (iii) the importer; (f) the reason for the recall, the nature of the defectiveness or potential defectiveness of the device and the date on which and the circumstances under which the defectiveness or potential defectiveness was discovered; and (g) a preliminary evaluation of the risk associated with the defectiveness or potential defectiveness of the device.
Read the Medical Devices Regulations on the Justice Laws Website ↗

Made under Food and Drugs Act, R.S.C. 1985, c. F-27.