Specifications Relating to Non-automatic Weighing Devices (1998), SI/98-81
Section 8
Cite as Specifications Relating to Non-automatic Weighing Devices (1998), SI/98-81, s. 8 · current to July 24, 2025 · regulation last amended June 17, 2019
Statute Watch
Get an alert when a bill in Parliament would change this section, and every time it moves.
The text
(1) The in-service limits of error set out in column I of the table to this section apply, in respect of each class of non-automatic weighing devices set out in column II of that table, according to the specified load.
(2) In the table to this section, m means the weight of the load placed on the weighing device, and is expressed in the number of verification scale intervals.
(3) The in-service limits of error that apply to a Class Oval outline with the text IIIHD inside weighing device are ±1 e where the load is not less than zero and not more than 500 verification scale intervals, and where the load exceeds 500 verification scale intervals, calculated by adding ±1 e for each additional 800 verification scale intervals or less, to a maximum of ±22 e.TABLEColumn IColumn IILimits of errorClass Oval outline with the text I insideClass Oval outline with the text II insideClass Oval outline with the text III insideClass Oval outline with the text IIII inside± 1 e0 ≤m≤ 50 0000 ≤m≤ 5 0000 ≤m≤ 5000 ≤m≤ 50± 2 e50 000 <m≤ 200 0005 000 <m≤ 20 000500 <m≤ 2 00050 <m≤ 200± 3 em> 200 00020 000 <m≤ 100 0002 000 <m≤ 4 000200 <m≤ 400± 5 e– – –– – –4 000 <m≤ 10 000400 <m≤ 1 200
Read the Specifications Relating to Non-automatic Weighing Devices (1998) on the Justice Laws Website ↗
Made under Weights and Measures Act, R.S.C. 1985, c. W-6.