Specifications Relating to Non-automatic Weighing Devices (1998), SI/98-81
Section 16
Cite as Specifications Relating to Non-automatic Weighing Devices (1998), SI/98-81, s. 16 · current to July 24, 2025 · regulation last amended June 17, 2019
Statute Watch
Get an alert when a bill in Parliament would change this section, and every time it moves.
The text
(1) Under stable environmental conditions, the deviation from zero as soon as the weight indication has stabilized following the removal of a load that has remained on a weighing and load-receiving element for up to 30 minutes shall not exceed
(a) in the case of a single range weighing device, one-half of the verification scale interval;
(b) in the case of a multiple range device, one-half of the verification scale interval for the range being tested; and
(c) in the case of a multi-interval device, one-half of the smallest verification scale interval.
(2) In the case of a multiple range device, following the removal of a load greater than the maximum capacity of the first range, which load has remained on the weighing and load-receiving element for up to 30 minutes, and immediately after switching the device to its lowest weighing range, the indication near zero shall not vary by more than one verification scale interval of the lowest weighing range over a period of five minutes after switching the device to its lowest weighing range.
Read the Specifications Relating to Non-automatic Weighing Devices (1998) on the Justice Laws Website ↗
Made under Weights and Measures Act, R.S.C. 1985, c. W-6.