Electronic Registers and Ancillary Equipment Incorporated in Metering Assemblies Specifications, SI/90-156
Section 29
Cite as Electronic Registers and Ancillary Equipment Incorporated in Metering Assemblies Specifications, SI/90-156, s. 29 · current to July 24, 2025 · regulation last amended March 22, 2006
Statute Watch
Get an alert when a bill in Parliament would change this section, and every time it moves.
The text
A register and all ancillary equipment shall remain within the applicable limits of error when tested under the following conditions:
(a) an ambient temperature in the range of -30°C to +40°C, except that a register and all ancillary equipment that are designed for use in a temperature-controlled environment shall be tested over the range of ambient temperatures that is specified by the manufacturer and marked on the name-plate of the register; and
(b) a relative humidity in the range of 10 to 95 per cent.
Read the Electronic Registers and Ancillary Equipment Incorporated in Metering Assemblies Specifications on the Justice Laws Website ↗
Made under Weights and Measures Act, R.S.C. 1985, c. W-6.